The phase fan opens, a 180° pulse mirrors it, and the fan closes again. Repeat the
refocusing and the echo peaks trace out T2 itself.
Ready.
Each thin arrow is one isochromat: a group of spins sharing a Larmor frequency. The
thick red arrow is their sum.
|signal|T2 envelopeecho peaks
Every echo peak lands on the T2 curve. Fitting those peaks is how a
multi-echo sequence measures T2.
Sequence and tissue
The first 180° comes at TE/2; echoes form at TE, 2TE, 3TE …
A CPMG train: one excitation, many refocusing pulses.
Sets T2*: how fast the fan opens between echoes.
The true decay the echo peaks are chasing.
More isochromats give a smoother, more realistic decay.
Simulated milliseconds per second.
|signal|0.000
echoes so far0
last echo--
predicted e−TE/T2--
t0 ms
What this figure is for
This is the figure to draw on the board when someone asks why we bother with a 180° pulse. The fan
on the left is the whole argument: dephasing is ordering, not loss, and order can be
reversed.
1/T2* = 1/T2 + γΔB0 → echo(n·TE) = M0 e−n·TE/T2
Run the train and watch the peaks, not the wiggles. The signal between echoes is
T2*-limited and tells you mostly about your shim. The peaks are T2-limited
and tell you about the tissue. A multi-echo sequence fits an exponential through those red dots,
which is exactly how a T2 map is made.
Press FID only. With no refocusing you see the raw T2* decay, and the signal is
gone long before T2 has run its course. The gap between that curve and the dashed
envelope is what the 180° pulses buy back.
The inhomogeneity slider does not move the peaks. Push it to 20 Hz: the fan opens faster,
the signal between echoes vanishes sooner, and the echo peaks stay exactly where they were.
That independence is the entire point of the spin echo, and it is why spin-echo T2
values are reproducible across scanners while T2* values are not.
Why TE controls contrast. Set T2 to 300 ms and then 1500 ms, and compare the
height of the last echo. Long-T2 tissue such as CSF still has signal at a long TE
while short-T2 tissue has none. That difference is T2 weighting.
Careful with the cartoon: the isochromats here keep a constant frequency, so refocusing is
perfect. Real spins diffuse between regions of different field and molecular tumbling randomises
their phase. That randomness is precisely T2, and it is what stops the echo train from
going on forever.