Dephasing & Rephasing Across the Sample

Looking straight down the bore at a slice of tissue. Each cell precesses at the rate its own local field dictates, and the coil adds them all up.

Press 90° excite to tip everything into the transverse plane.

Cell shading is the field offset at that location; arrow direction and colour are the phase accumulated there.

The same spins redrawn from one origin, with the vector sum in red.

|signal| T2 envelope

The field map

A smooth shim error straight across the slice.
A sinus, an ear canal or a metal clip: a big offset over a small region.
Random molecular-scale dephasing. No pulse can undo this.
The echo lands at TE = 2τ after the 90°.
Number of cells across the slice.
Simulated milliseconds per second.
|signal|0.000
T2 envelope1.000
field range0 Hz
t since 90°0 ms

What this figure is for

The previous figures treat the sample as a single vector. This one insists on the fact that position matters: every cell sits in a slightly different magnetic field, so every cell accumulates phase at a different rate. The coil has no idea which cell is which. It receives one number, the vector sum, and that number collapses as soon as the cells disagree.

Try this in class: set τ to 500 ms, press Run a spin echo, and ask the room to predict where the echo peak will fall before it happens. Then change T2 and ask them to predict how tall it will be.