IDNE 701: Neuroengineering / Syllabus
Mid-Semester Milestone · Week 7

Worksheet 7.1: Preliminary Design Review (PDR) & Evaluation Rubric

Formal design milestone assessment for student neurotechnologies. Evaluates biophysical contrast justification, sensor hardware architecture, noise budget margin, signal encoding/reconstruction strategy, and Grand Challenge 06 (Neonatal & Pediatric Brain Monitoring) alignment.

01.

Project Identification & Review Mode

👶 Grand Challenge 06 Core Criteria Projects focused on neonatal and pediatric neuromonitoring must explicitly defend mechanical compliance (skin pressure < 20 mmHg, no abrasion), ambient NICU rejection (phototherapy lighting, incubator radiant heat), and biomarker clinical validity (e.g., FTOE, pulsatility index, burst suppression).
02.

Structured Evaluation Matrix (100 Points Total)

Evaluate the presenting team across 5 core engineering domains (20 points each).

Evaluation Domain Criteria & Technical Benchmarks Score (0–20) Weighted Pts
1. Biophysical Contrast Mechanism
Weeks 2–4 Material
• Rigorous forward physical model linking neural state to perturbation field (Maxwell, Bloch, Diffusion, or Bioheat).
• Clear physical justification for chosen contrast mechanism (e.g., BOLD, phase velocity, optical density).
• Identification of physical limits on spatial and temporal resolution.
18 / 20
2. Sensor Array & Noise Budget
Week 5 Material (Worksheet 5.1)
• Calculated Johnson-Nyquist thermal limits and sample-noise crossover ($Q_U / Q_L > 2.5$).
• Transducer array geometry and inter-element decoupling strategy (preamp decoupling, overlap, active shield).
• Analog front-end (AFE) amplification, noise figure ($NF < 1.5\,\text{dB}$), and dynamic range specification.
18 / 20
3. Signal Encoding & Inversion Plan
Weeks 6 & 8 Material
• Mathematical formulation of the inverse problem (e.g., MBLL matrix, k-space trajectories, lead-field inversion).
• Inversion stability, condition number ($\kappa$), and ill-posedness mitigation (regularization, Tikhonov).
• Defined extraction pipeline for clinical biomarkers (e.g., $\text{rSO}_2$, $\text{FTOE}$, CBV, latency).
17 / 20
4. Clinical Translation & NICU Fit
Grand Challenge 06 Alignment
• Specific relevance to bedside neonatal monitoring (HIE, IVH, or seizures).
• Mechanical safety: Contact pressure < 20 mmHg, no scalp abrasive prep, biocompatible substrate.
• Environmental robustness: Phototherapy optical filtering ($\sim 450\,\text{nm}$), incubator thermal limits ($< 41^\circ\text{C}$).
19 / 20
5. Presentation, Visuals & Concept Boards
Design Review Delivery
• 12-minute pitch clarity, professional system architecture block diagrams.
• Effective concept pin-up boards visualizing physical device ergonomics and patient montage.
• Rigorous and candid defense during 8-minute panel Q&A.
18 / 20
Total Milestone PDR Score:
Threshold for Unconditional Phase 4 Advance: 80 / 100
90 / 100
03.

Qualitative Critique & Action Items

What is most technically compelling about the chosen transducer, contrast physics, or biomarker pipeline?
Where is the design most fragile? (e.g., motion artifact vulnerability, ill-conditioned inverse matrix, thermal heating, excessive skin contact pressure).
Specific technical revisions required prior to virtual simulation lock-in.
04.

Milestone Verdict & Phase 4 Advancement Sign-Off