Positron Emission Tomography · Radon Transform, Coincidence LORs, Filtered Backprojection (FBP) & Iterative MLEM with Attenuation Correction
In 2D PET, each line integral across tracer activity \(f(x, y)\) is parameterized by normal angle \(\theta \in [0, \pi]\) and signed radial displacement \(s\) from the scanner isocenter:
Projecting sinogram values back along each LOR without filtering accumulates the image, but introduces an intrinsic \(1/r\) point-spread blurring kernel (\(f_{\text{unfiltered}} = f \ast \frac{1}{r}\)). In Fourier space, the Fourier Slice Theorem proves that 1D projections correspond to radial slices through 2D k-space. The Ram-Lak ramp filter (\(|k|\)) compensates exactly for the high density of radial sampling near the origin. Apodizing with a Hann window (\(H(k) = 0.5 + 0.5\cos(\pi k / k_{\text{max}})\)) prevents high-frequency Poisson noise blow-up.
Filtered Backprojection assumes deterministic, continuous line integrals and easily produces negative artifactual pixel values in low-count regions. Maximum Likelihood Expectation Maximization (MLEM) models the discrete Poisson statistics of radioactive decay:
Before escaping the patient's body, 511 keV photons can undergo Compton scattering or photoelectric absorption. The survival probability along an LOR is: