PET Sinogram & Tomographic Reconstruction Workbench

Positron Emission Tomography · Radon Transform, Coincidence LORs, Filtered Backprojection (FBP) & Iterative MLEM with Attenuation Correction

1. Emission Phantom Click to add Hot Tumor
2. Measured Sinogram p(s, θ) s × θ (Radon Transform)
3. Reconstructed Activity Filtered Backprojection
Reconstruction Algorithm:
Tracer Dose & Count Statistics High (Low Noise)
Low Dose (High Poisson Noise) High Dose (Clinical)

Mathematical Foundations of PET Tomography

1. The Radon Transform & Sinogram Geometry:

In 2D PET, each line integral across tracer activity \(f(x, y)\) is parameterized by normal angle \(\theta \in [0, \pi]\) and signed radial displacement \(s\) from the scanner isocenter:

\[ p(s, \theta) = \iint f(x, y) \, \delta(x \cos\theta + y \sin\theta - s) \, dx \, dy \]
A single isolated hot tumor at polar coordinate \((r_0, \phi_0)\) traces a pure sine wave across the Radon plane: \(s(\theta) = r_0 \sin(\theta - \phi_0)\). The sinogram is the collection of all such intersecting sinusoids.

2. Unfiltered vs. Filtered Backprojection:

Projecting sinogram values back along each LOR without filtering accumulates the image, but introduces an intrinsic \(1/r\) point-spread blurring kernel (\(f_{\text{unfiltered}} = f \ast \frac{1}{r}\)). In Fourier space, the Fourier Slice Theorem proves that 1D projections correspond to radial slices through 2D k-space. The Ram-Lak ramp filter (\(|k|\)) compensates exactly for the high density of radial sampling near the origin. Apodizing with a Hann window (\(H(k) = 0.5 + 0.5\cos(\pi k / k_{\text{max}})\)) prevents high-frequency Poisson noise blow-up.

3. Iterative MLEM Reconstruction:

Filtered Backprojection assumes deterministic, continuous line integrals and easily produces negative artifactual pixel values in low-count regions. Maximum Likelihood Expectation Maximization (MLEM) models the discrete Poisson statistics of radioactive decay:

\[ \hat{f}_j^{(n+1)} = \hat{f}_j^{(n)} \frac{1}{\sum_i c_{ij}} \sum_i c_{ij} \frac{p_i}{\sum_k c_{ik} \hat{f}_k^{(n)}} \]
Each iteration compares projected estimates against actual measured counts, guaranteeing strictly non-negative activity and superior signal-to-noise ratios.

4. 511 keV Photon Attenuation Correction:

Before escaping the patient's body, 511 keV photons can undergo Compton scattering or photoelectric absorption. The survival probability along an LOR is:

\[ A_i = \exp\left(-\int \mu(x,y) \, dl\right) \]
independent of where along the LOR the emission took place! Without attenuation correction (CT or MRI-based \(\mu\)-map), deep brain structures appear artificially hypometabolic while the cortical periphery creates false hypermetabolic rims.